|
www.MathResolutions.comSoftware Products for the Radiological Sciences |
|
| Home Page | Product Review | Program Manuals | Download Programs | Purchase | Site Map |
| Dosimetry Check | MarkRT (VGRT) | RtDosePlan | System 2100 | MillComp | C++ Library |
To compute the dose to the patient from a treatment portal,
pencils are first created that cover the treatment field and patient.
The area of a beam field is simply divided up into small pixels
(in a plane perpendicular to the beam), each pixel being a separate
pencil. Whether a pencil beam algorithm or a full superposition
algorithm, it is necessary to trace each pencil through the patient
as illustrated in the below image where we have lighted up every
other pencil.
Typically the density of the
material the beam is going through is noted and one computes terma along
each ray. However, to compute the dose to the patient
one must know the intensity of radiation that reaches each ray.
In a planning system to know the intensity that is reaching each pencil
on the surface
of the patient, a source model is used that can compute the intensity
down stream from the collimation system on any point on a plane
perpendicular to the central ray, as illustrated in the below
image.
The source model must take into account everything, the flattening
filter, the collimator system, the multi-leaf system, wedges, and
any dynamic movements of those things to arrive at the final intensity
that reaches each pencil.
Dosimetry Check does not have a source model, which is the
whole point of Dosimetry Check. In Dosimetry Check, the
source model is replaced with a measured source model.
If you were to put a piece of x-ray film on the couch and expose to
the treatment portal straight down for the whole beam on time, you will
have a picture of the field intensity that takes into account everything,
such as leaf leakage, since you have just measured it all. In Dosimetry Check
you can use the EPID or a diode or ion chamber array for greater
convenience. The
distance to the patient surface for each pencil is known from the CT scans
and beam position.
The beam film (taken BEFORE or without the patient)
gets darker the more radiation a spot gets.
As noted above, the port image without the patient in the beam is a
record of how much radiation each pixel gets.
It is necessary to convert this information to a unit that will
make it possible to compute the dose in centiGray to the patient
from each pencil. Each spot on the
image is mapped to the monitor units that would produce the same darkness
at the center of a 10x10 cm field size (i.e. the field size that you
calibrate the accelerator to).
We call this resultant number the "relative monitor units" (RMU).
Shown below is an EPID image of a modulated field for a head
and neck case after mapping to RMU units (where white is more radiation).
To accomplish this mapping you must run a calibration curve of
the pixel value at the center of a 10x10 cm field versus
monitor units.
This curve is used to map the
pivel values on the beam images above to RMU.
However, for an EPID, ion chamber or diode array, integration is a
linear process with zero intercept.
So one can use a single measurement of a 10x10 cm
field to determine the same calibration line. This means
only one 10x10 field need be measured with a clinical case
to map the fluence images to RMU values. But an EPID or similar
device may generate internal scatter within the device. This
is corrected for by doing a deconvolution with the inverse
of the point spread function of the device to arrive at in
air fluence in RMU.
Once converted to RMU, Dosimetry Check can compute the dose from a
beam to the patient in centiGray.
Here, zero RMU is black, whiter is larger RMU.
The pixels on the RMU image are used as "weights" for the corresponding
pencil beams. The fluence map that we have derived here from
a calibrated image of the beam completely determines the dose to the
patient. This process is similiar to how a planning system computes
the dose, except here we are using the measured fluence instead of computing
the same from knowledge and models of what kind of things modulate the beam.
By starting with measurement instead of a model, we verify the dose
and dose distribution that the patient receives.
The dose computed by Dosimetry Check may be compared directly to that
computed by the planning system.
ADAC IMRT plan (green) versus Dosimetry Check (magenta) with inhomogeneity
corrections on, courtsey of Dr. Tianyou Xue.